ADM 6725 E10
KA
T
8 - 3
Measuring Data Transfer to PC
8.3
Optional Parameters for MOVEFAST and
FLXtoTXT
The listed formatting options can be used independently. If a formatting option is not
specified, the programs will use the default setting.
/Cx
Specifies the serial port
for MOVEFAST only. If this parameter is not specified,
COM2
will be used. /Cx has no effect for program FLXtoTXT
.
/C1=COM1 ... /C4=COM4
/N
An existing file will be overwritten without asking for confirmation.
/ONLY=n
Output of measuring data set number n only. In case n is negative, the counting
starts from the end.
Without this parameter, all measuring data sets will be output.
/ONLY=3
output of measuring data set No. 3
/ONLY=
-
2
output of measuring data set which was stored before the last
one
/FP=n
Every numerical value will be represented by n characters. (Default=9)
/FP=9
results in -1,00E+05 for -100380,37
/FP=10
results in t -100380,37 for -100380,37
/
,
Forces a decimal comma (=default)
Example: 7,13
/
.
Forces a decimal point
Example: 7.13
/ST
+
Separates measured values, total counts, etc. by a tabulator. (Default).
7,13
⇒
⇒
⇒
⇒
+3,22
⇒
⇒
⇒
⇒
-0,44
⇒
⇒
⇒
⇒
/ST-
The tabulator is not used for the separation of columns.
/SS
+
Separates measured values, total counts, etc. by a semicolon.
7,13; +3,22; -0,44;
/SS-
The semicolon is not used for the separation of columns. (Default)
/TRIM
The rows with measured data will not contain any space characters.
7,13;+3,22;-0,44;
/VL
+
The flow velocity, which the measured value was calculated from, will be additionally
output in the first column ('VELOC') in [m/s].
0,421;7,13;+3,22;-0,44;
Flow Velocity=0,421 m/s
/M2
+
Even when only one total count value is being stored in memory, 2 columns will be
generated in the output file ('Q_POS','Q_NEG'). (Default)
/M2
-
If during measurement the forward and return total quantities have never been stored
simultaneously, only one column will be generated ('QUANT'). Positive values belong
to the forward total quantity, negative values belong to the return total quantity.
/MM
+
Always generates two columns for forward and return total quantities ('Q_POS',
'Q_NEG'). If no total quantity counts were stored in the row of measured values, the
columns are kept empty. /M2+- is ineffective.
/MM
-
Columns for forward and return total quantities are always omitted. /M2+- is
ineffective.
/2T
+
Even when only one temperature was stored, two columns ('T1','T2') will be
generated.
/2T
-
When only one temperature was stored, only one column will be generated. (Default)
/D
"+
Limits every line of the parameter record with quotation marks.
"pipe material : copper"
/D
"
-
Does not limit lines of the parameter record with quotation marks. (Default)
pipe material : copper
/HD
+
Generates column titles for every measuring data set. (Default)
\DATA
\* MEASURE; Q_POS; Q_NEG; T1; T2;
7,13; +3,22; -0,44; 20,3; 40,8;
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...