7 - 8
ADM 6725 E10
KA
T
Program Branch SPECIAL FUNCTIONS
7.5
Selection of the Measuring Rate
Here, you need to decide, through selection of the measuring rate, upon the power
consumption of the instrument.
•
If you select the measuring rate
MAX
, measurements are carried out
continuously.
•
If you select the measuring rate
NORMAL
, the measuring rate is equivalent to
about 20 % of the maximum rate. This means, that in this mode the power
consumption of the instrument is 80 % less than with the device set for the
maximum measuring rate. This set-up would be an essential aspect when
FLUXUS is operated in battery mode and needs to measure over long periods.
Special Funct.
"
Select Meas.Rate
S8
Here, this display of program branch SPE-
CIAL FUNCTIONS requests you to decide
on the continuity of the measurement and
subsequently upon the power
consumption of the flowmeter.
>NORMAL< max
Measuring Rate:
S8-1 Make your selection with regard to the
measuring rate.
Confirm your choice by pressing
ENTER.
!
MAIN MENU
7.6 Battery
Charging
Special Funct.
"
Charge Battery
S9
With this function, the battery charging
procedure can be activated.
15.0 h
Charg.time Batt.
S9-1 Enter the desired charging time of the
battery up to a maximum of 15 h.
Note!
The above display does not appear if the battery charging process is already running
in the background. In such a case, the following message appears:
(15:00) 12:30 *
Battery Charging
S9-2
(15:00)
shows the charging time that had been entered.
12:30
indicates the remaining charging time (count-down).
*
signals in second intervals that charging process is running.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...