Wall Thickness Measurement
Measuring-WTM
par>MEA<opt sf
M* Select from the MAIN MENU the program
branch MEASURING.
NO DATA !
par>MEA<opt sf
F4 If this error message appears:
•
you did not enter the required input
parameters completely;
•
or, you did not change the set 0.0 m/s
for a unknown material to its correct
sound velocity.
12.5.1 Measurement of the Wall Thickness
mm?
Wall thickness
M* This display appears after calibration, if wall
thickness was selected as the quantity of
measurement. When there is no valid
measuring value, the unit of measurement
and a question mark are shown on the
lower display line.
3.51 mm
Wall Thickness
√
M* Firmly press the probe bearing a film of the
acoustic coupling component against the
test piece (the pipe) for the measurement of
its wall thickness. As soon as a valid
measuring value has been obtained, a tick
sign and the thickness will be shown.
Note:
The actual sound velocity of a material often differs substantially from the values
published in the literature.
To minimise errors during the measurement
of the wall thickness:
•
Adjust the values stored in FLUXUS
according to the actual sound velocity of
the test piece, using a comparative block
of known dimensions of the same
material.
•
The comparative block should be even
and plain.
•
The thickness of the comparative block
should be equal to the maximum
thickness of the test piece (the pipe).
•
Enter the thickness of the comparative
block in program branch PARAMETER.
(See chapter 12.4.2)
.
•
Measure the sound velocity of the
comparative block.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...