2 - 6
FLUXUS ADM 6725 E10
KA
T
Description of Flowmeter
Attention!
If the key combination
has been initiated, the display appears in the
factory pre-set language version again.
2.3.3 HotCodes
HotCodes are key sequences in order to call up specific operational modes/set-
ups of FLUXUS.
(The previous chapter dealt with the key sequences, which form the so-called HotCode, to select a
specific language.)
HotCodes can be entered, when FLUXUS displays the MAIN MENU after
switching ON.
The following table gives an overview of the most important HotCodes:
HotCode See
chap-ter
Mode/set-up
Will this mode
remain activated
after switching the
unit OFF?
Will this mode
remain activated
after a cold start
(INIT)?
How to de-activate the
mode/set-up done by
HotCode
9 0 9 0 4 4
9 0 9 0 4 9
2.3.2 Display
language:
English
German
(other language codes,
see chapter 2.3.2)
Yes No,
the factory pre-set
display language
appears again
(default language).
Enter the code of another
language.
0 7 1 0 4 9
13.1
Activation of Super User
Mode
No
No
Switch FLUXUS OFF.
2 3 2 8 8 8
14.5
Change of serial protocol
for graphics printing
Yes
Yes
Enter the same code.
0 0 7 0 0 1
19.1
Enable or disable time
programmable
measuring mode
Yes
Yes
Enter the same code.
0 0 7 0 0 2
15.4
Enable or disable offset
correction facility for heat
quantity measurement
Yes
Yes
Enter the same code.
0 0 7 0 0 3
5.4.2
Change of data storage
mode for data logging of
measured values
Yes
Yes
Enter the same code.
0 0 7 0 0 4
5.4.3
Change of data storage
mode for total quantity
counters
Yes
Yes
Enter the same code.
0 0 7 0 0 5
5.4.1
Change profile
correction for 'Flow
velocity' measurement
Yes
Yes
Enter the same code.
0 0 7 0 0 8
11.3
Basic unit selection for
the measurement of heat
quantity
Yes
Yes
Enter the same code.
0 0 7 0 0 9
5.2.1
Definitions for
description of measuring
point
Yes
Yes
Enter the same code.
0 7 1 0 0 0
4.3.5
Change of parameter
limit for transducers of
type ‘Q’
(Set-up of low limit for inner pipe
diameter)
Yes
Yes
Enter the same code.
0 7 1 0 0 0
5.3.3
Set-up of limit for cut-off
flow velocity
Yes
Yes
Enter the same code.
2 3 2 0 0 0
6.6
Set-up for data format of
serial output
Yes
Yes
Enter the same code.
0 0 6 9 9 9
See next
page
Sequential display of
inputs, which can be
called by HotCodes
separately
No
Yes
Switch FLUXUS OFF.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...