Wall Thickness Measuring
12 Wall Thickness Measurement (WTM)
12.1 Overview of the WTM Option
When equipped with the WTM Option,
FLUXUS
can operate as a wall thickness
gauge. For this, a dedicated wall thickness sensor, which plugs directly into the
sensor connection port, is supplied. FLUXUS is able to automatically recognise
that the wall thickness sensor has been connected to the instrument. The value
measured through the wall thickness function can easily be transferred into the
current parameter record for the flow measurement, and may also be protocolled if
necessary. The determination of the longitudinal sound velocity of a material is
also possible.
FLUXUS
with WTM option uses a modified transit-time method to determine the
thickness or sound velocity of a material (test piece). The sensor probe emits a
short ultrasonic pulse into the test piece. The pulse signal is reflected by the
boundary layer of the test piece and received again by the sensor probe. The time
difference between the emitted and received signal is a measure for the thickness
of the test piece
(with known sound velocity of the material)
, or for its longitudinal sound velocity
(with known thickness of the material)
.
With the model
FLUXUS ADM 6725
no calibration
(zero adjustment for the probe)
is
necessary.
Probe cable
With plug
!
for FLUXUS
socket and
with plug
!
for probe
socket
Wall thickness sensor assembly
"
Sensor probe with
connection socket
on the side for
probe cable
Measuring principle
Sensor probe
Probe cable
Boundary layer of test piece
12.2 Instrument
RESET
INIT
3x OFF
ON
#
Connection port for flow
transducers channel A or wall
thickness sensor
#
Connection port for flow
transducers channel B or wall
thickness sensor
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...