ADM 6725 E10
KA
T
6 - 3
Program Branch OUTPUT OPTIONS
The MAIN MENU appears
again. The program branch
MEA is already highlighted.
6.2 Selection of the Quantity and the Unit of
Measurement
There is a choice of the following quantities of measurement available:
•
Volume flow
•
Mass flow
•
Flow velocity
•
Fluid sound velocity
FLUXUS measures the flow velocity in the pipe directly. FLUXUS calculates the volume
flow by multiplying the flow velocity with the cross-sectional area of the pipe. The mass
flow is calculated by the volume flow with the density of the medium.
Output Options
par mea >OPT< sf
O1
Select and confirm from the MAIN MENU the
program branch
OPT
. The page for the
selection of the quantity of measurement will
appear next.
Meas. Quantity
"
Volume Flow
O3
Select the desired quantity of measurement.
Confirm by pressing ENTER.
Flow Velocity
O4
Mass Flow
O5
Fluid velocity
O6
Heat flow
O7
(optional)
6.2.1 Quantity of Measurement: Volume Flow
With the quantity of measurement
volume flow
selected, the next display will show 'Volume
flow in:'.
Note:
The unit, which has been selected previously, will be shown in the scroll list first.
Volume in:
"
m3/h
O3-1
Scroll through the list and select your desired
unit of measurement.
m3/min O3-2
m3/s O3-3
l/h O3-4
l/min O3-5
l/s O3-6
USgph O3-7
USgph = US gallons/hour
bls/d O3-8
bls/d = barrels/day
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...