Process Interfaces
13 - 4
ADM 6725 E10
KA
T
Note:
The entered full-scale value is valid for all activated analogue process interfaces
(current loop, voltage output, frequency output).
13.2 Set-up of Pulse Output
Note:
The display 'Pulse Output'
only
appears in program branch OPT
if
FLUXUS is
equipped with a binary output, which has been installed as pulse output (source
item: Impulse) in program branch SPECIAL FUNCTIONS (see chapter 7.7.5).
When the flow velocity (display O4) is selected as the quantity of measurement, the use of
the pulse output is not possible (because it is technically not sensible).
no >YES<
Pulse Output
O16
Select
YES
, if you wish
to activate the pulse
output.
0.01 m3
Pulse Value
O17
The display for the input of the pulse value
appears.
FLUXUS displays the unit of volume or mass (
m3
in our
example) according to the previously selected unit of
measurement (in this case: m3/h acc. to display O3-1). The
pulse value is the quantity of flow volume or mass per
transmitted pulse.
The display 'Pulse Width' appears next.
The pulse width to be set depends on certain
technical characteristics of the device (e.g. counter, totaliser, PLC), which will be
connected with the pulse output:
30 ms
Pulse Width
O18
Enter the 'Pulse Width' of the pulse value
between 30 and 1000 milliseconds.
Now, FLUXUS will give information regarding the maximum possible flow in the pipe that the
pulse output can work with. This is calculated from the data given for pulse value and pulse
width:
31.3 m3/h
INFO: Max-Value
O19
ENTER confirm
Attention!
If the
actual
flow exceeds this 'Max-Value', the pulse output will not function
correctly. In such a case, the pulse value and pulse width should be changed to
accommodate the flow conditions.
13.3 Set-up of Alarm Outputs
Note:
The display 'Alarm Output'
only
appears in program branch OPT
if
FLUXUS is
equipped with a binary output, which has been installed as alarm output (source
item: Limit) in program branch SPECIAL FUNCTIONS (see chapter 7.7.5).
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...