5 - 14
ADM 6725 E10
KA
T
Program Branch MEASURING
+
O
Press this key for information about total
volume quantity for
positive flow
direction.
-
O
Press this key for information about total
volume quantity for
negative flow
direction.
Note:
•
To show the totalised throughputs for a
certain measuring channel
, this
channel must be shown on the display.
•
The status indicator
S IG N A L
(red/green) and keypad operations influencing
the measuring mode are linked to the measuring channel, which is currently
shown on the display.
Reset
the two flow totalisers
to zero
:
Press key
O N
O
.
Deactivate
flow totalising:
Press key
O F F
O
.
5.3.5 Disturbances during Measurement
A
question mark
on the bottom line at the right hand side of the display will indicate
signal loss. The pipe cannot be sufficiently acoustically penetrated and the received
signal cannot be used for measurement. The last correct reading will remain on the
display:
54.5 m3/h?
A: Volume Flow
M8
E NT E R
O N
O
Return to the bar graph display.
Activates flow totalisers.
Suggested actions to overcome the problem
FIRST:
Wait a little while until the acoustic
contact is established again.
Temporarily, there might be a higher
proportion of gaseous or solid particles
in the flowing medium.
SECOND:
Check that there is a film of acoustic coupling
component between the transducers and
pipewall. Renew film if necessary. Start the
sequence to position the transducers again.
Adjust transducers for maximum acoustic
contact.
THIRD:
Change from using the reflection to the
diagonal-path mode. The signal
attenuation might be too high because
of high fluid viscosity or deposits on the
inner pipewall.
FOURTH:
Find a more suitable measuring point along
the pipework.
(See chapter 3)
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...