ADM 6725 E10
KA
T
4 - 3
Program Branch PARAMETER
4.2 Input of Pipe Parameters
Parameter
>PAR< mea opt
sf
P1
From the MAIN MENU select and confirm
the program branch PAR.
for Channel A:
Parameter
"
P2
(Here, the measuring channels can be
selected, max. A ... H, for devices with
integrated measuring point multiplexer.)
"
select
ENTER confirm
100.0 mm
Outer diameter
P3
This display of the program branch PAR
appears next.
(The previously entered outside diameter
will be displayed.)
The next display will request the entry of the pipe's outside diameter.
4.2.1 Outside Pipe Diameter / Circumference
100.0 mm
Outer Diameter
P3
(The previously entered outer diameter will be displayed.)
Enter here the outside diameter of the pipe
within the range of the transducer
characteristics.
In program branch SPECIAL FUNCTIONS, you may enable the set-up facility for
the pipe circumference
(see chapter 7.7.2)
.
Additionally by HotCode definition, you can select the entry mode and make it
cold-start resistant.
(See also chapter 2.4.4)
+
O
15 2
15 2
15 2
Immediately after pushing the BRK key,
enter this key sequence from the MAIN
MENU.
off >0N<
Pipe Circumfer.
This display appears after the facility was
enabled.
(In case of using the HotCode, it appears within a
display group.)
Confirm
>ON<.
Entering 0 (zero) in display
‘Outer Diameter’, FLUXUS switches to the
circumference entry facility.
If selecting
>OFF<,
FLUXUS displays an
error message, if 0 (zero) is entered in
display ‘Outer Diameter’.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...