All stems are for use WITHOUT cement, long term fixation being achieved through bone
ingrowth and union of the coating and the host bone. Stems are available in various sizes to
accommodate anatomical variations of the femur.
IMPORTANT: Adjunctive proximal fixation / support is
required
for Size 10, 11 and 12 monobloc,
11mm and 12mm diameter modular 250mm length stems and 13mm diameter 300 length
modular stems and is
recommended
for all the Securus and Furlong® H-A.C. revision stems,
both monobloc and modular. Where there is loss of proximal bone stock, or poor proximal bone
quality, bone grafting or other adjunctive proximal fixation / support is advised for implant
stability. It is important that the prosthesis is not distally fixed without proximal support.
A range of Furlong® H-A.C. Acetabular Cups
are available for use with this system, refer to
155-019.
Femoral Heads
of various diameters and neck lengths are available for use with this system in
Cobalt Chrome alloy or ceramic. Ceramic heads are made from alumina ceramic or Biolox delta.
Special Instructions For Use are required for ceramic heads refer to 155-020. JRI recommend
the use of highly crosslinked UHMWPE (CLP-75) with larger metal head diameters because of
reduced volumetric wear.
Bone Substitute granules, blocks and pastes
are available for use with this system, please
refer to 155-028.
All types of bone screw are available in a variety of lengths and are manufactured from Titanium
alloy Ti-6Al-4V, as are the modular revision connection screw and blanking plug.
Smaller sized implants are intended for patients with smaller bone and regular Body Mass Index
and could be inappropriate for other patients.
Note:
Components should only be used with other compatible components of the Securus and
Furlong® H-A.C THR Systems, with the corresponding taper connections. Implant components
from one manufacturer MUST NOT be used together with those of another manufacturer,
since compatibility of mating parts cannot be assured
except for product combinations
specifically tested and approved by JRI
.