7 SETUP/MAINTENANCE
Invacare® Formula™ CG Seating
80
Part No 1143155
7.5
Adjusting/Replacing Telescoping Front Rigging Support
Adjusting
1.
Remove the two cap screws, spacers and threaded blocks securing the telescoping front tube to the side rail.
2.
Slide existing telescoping front rigging support to one of six depth positions.
3.
Use the existing two cap screws, spacers and threaded blocks to secure the telescoping front tube to the side rail:
4.
Repeat STEPS 1 to 3 on the opposite side, if desired.
Replacing
1.
Remove the two cap screws, spacers and threaded blocks securing the telescoping front tube to the side rail.
2.
Remove existing telescoping front rigging.
3.
Perform one of the following:
•
Reclining Seating Systems Only - Refer to Detail “A” in FIGURE
2 to determine the mounting position for the new telescoping front tubes.
•
Non-reclining Seating Systems - The telescoping front tubes can be positioned using one of six mounting positions. Proceed to STEP 4.
4.
Use the existing two cap screws, spacers and threaded blocks to secure the telescoping front tube to the side rail:
5.
Repeat STEPS 1 to 4 on the opposite side, if desired.
WARNING
If the telescoping tubes need to be extended greater than two inches, then the seat must be repositioned rearward to
ensure stability - otherwise personal injury and/or damage to the wheelchair and surrounding property may result.
For this procedure refer to FIGURE
2 on page 81.
Содержание Formula CG Seating
Страница 25: ...3 LABEL LOCATIONS Part No 1143155 25 Invacare Formula CG Seating 3 2 TDX SP with Vent Tray Option...
Страница 121: ...NOTES Part No 1143155 121 Invacare Formula CG Seating Notes...
Страница 122: ...NOTES Invacare Formula CG Seating 122 Part No 1143155 Notes...