8 TROUBLESHOOTING
Invacare® Formula™ CG Seating
120
Part No 1143155
Battery indicator flashes the charge level is
low - immediately after recharge.
Battery failure.
Malfunctioning battery charger.
Electrical malfunction.
Check batteries for shorted cell. Replace if necessary.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare for Service.
Battery indicator flashes the charge level is
low - too soon after being recharged.
Batteries not charged.
Weak batteries.
Have charger checked.
Replace batteries if necessary. Refer to Replacing the
Batteries and/or Battery Cables on page 83.
Motor “chatters” or runs irregular.
Motor/gearbox malfunction.
Stop use of Wheelchair. Contact Dealer/Invacare for Service.
Joystick erratic or does not respond as
desired.
Damaged motor coupling.
Electrical malfunction.
Controller programmed
improperly.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare for Service.
Contact Dealer/Invacare to have controller reprogrammed.
Wheelchair does not respond to commands. Electrical malfunction.
Contact Dealer/Invacare for Service.
Power indicator off - even after recharging.
Poor battery terminal connection.
Have terminals cleaned.
SYMPTOM
PROBABLE CAUSE
SOLUTIONS