8 TROUBLESHOOTING
Invacare® Formula™ CG Seating
116
Part No 1143155
SPJ™+, SPJ+ w/PSS or SPJ+ w/ACC Joysticks
The joystick information gauge and the service indicator give indications of the type of fault or error detected by the control module. When a fault is
detected, the wheelchair may stop and not drive. The LEDs on the information gauge may flash in a particular pattern or the service indicator light
will flash. The number or type of flashes indicates the nature of the error. If multiple errors are found, only the first error encountered by the control
module will be displayed.
Information Gauge Display Diagnostics
DISPLAY
DESCRIPTION
DEFINITION
COMMENTS
All LEDs are off.
Power is off.
All LEDs are on.
Power is on.
Fewer than three LEDs on
implies reduced battery charge.
Left RED LED is flashing.
Battery charge is low.
The batteries should be charged
as soon as possible.
Left to Right “chase” alternating
with steady display.
Joystick is in programming,
inhibit and/or charging mode.
The steady LEDs indicate the
current state of the battery
charge.
All LEDs are flashing slowly.
Joystick has detected Out-of-
Neutral-at-Power-Up mode.
Release the joystick back to
Neutral.
Information
Gauge Display