7 SETUP/MAINTENANCE
Invacare® Formula™ CG Seating
74
Part No 1143155
7 Setup/Maintenance
7.1
Setup/Delivery Inspection
Check all parts for shipping damage. In case of damage, DO NOT use.
Ensure clothing guards are secure
Ensure arms are secure but easy to release and adjustment levers engage properly.
Ensure adjustable height arms operate and lock securely.
Make sure all electrical connections are secure.
Make sure drive lock-out operates properly.
Make sure elevate systems drive with reduced speed when seat is in elevated position.
Check that cables are routed and secured properly to ensure that cables do NOT become entangled and damaged during normal operation of
seating system.
Ensure proper operation of powered functions (Example: seating and legrests).
Setup/delivery inspection should be performed by dealer at time of delivery/set up.
Initial adjustments should be made to suit your personal body structure needs and preference. Thereafter weekly, monthly
and periodic inspections should be performed by user/attendant between the six month service inspections. Refer to User/
Attendant Inspection Checklists on page 75
Every six months, and as necessary, take your wheelchair to a qualified technician for a thorough inspection and servicing.
Refer to Service Inspection on page 77.
Refer to wheelchair base owner’s manual for additional safety inspection and troubleshooting information.
Содержание Formula CG Seating
Страница 25: ...3 LABEL LOCATIONS Part No 1143155 25 Invacare Formula CG Seating 3 2 TDX SP with Vent Tray Option...
Страница 121: ...NOTES Part No 1143155 121 Invacare Formula CG Seating Notes...
Страница 122: ...NOTES Invacare Formula CG Seating 122 Part No 1143155 Notes...
Страница 123: ...NOTES Part No 1143155 123 Invacare Formula CG Seating Notes...