TM 11-6625-3071-14
3478A
Table 4-2. Recommended Test Equipment
Instrument
Critical Specification
Recommended Model
Use*
Digital Voltmeter
Range:30OmV to 300V
-hp- Model 3456A
PCT
Accuracy: + .004%(30V,300V)
± .005%(300mV,3V)
± .025%(30mV)
DC Voltage
Voltage:3OmV to 300V
Systron Donner
PCT
Standard
Accuracy: ± .005%
Model M107
AC Calibrator
Frequency:2OHz to 300KHz
Fluke Model 5200A
PCT
Voltage:30mV to 300V
and Model 5215A
Accuracy: .1 %
Voltage Stability: ± .02%
AC/DC Current
DC Current:100mA to 1A
Valhalla Model
PCT
Source
Accuracy: ± .03%
2500
AC Current: 100mA to 1A
Frequency: 1KHz to 5KHz
Accuracy: .1 9%
Resistance
Guideline Model:
Standard
Resistance: 10 ohm
9330/10 or
PC
Accuracy: i .0005%
9330A/10
Resistance: 100 ohm
9330/100 or
PC
Accuracy: f .0005%
9330A/100
Resistance:1K ohm
9330/1K or
PC
Accuracy: ± .0005%
9330A/1 K
Resistance: 1 OK ohm
9330/10K or
PC
Accuracy: ± .001%
9330A/10K
Resistance: 100K ohm
9330/100K or
PC
Accuracy: ± .001%
9330/100K
Resistance:1M
9330/1M
PC
Accuracy: ± .002%
Resistance: 10M
9330/10M
PC
Accuracy:
±
.01%
Desktop Computer
HP-IB Capability
-hp- Model 85,
CTO
9825,9826,9835,or 9845
Bus System
HP-IB Control Capability
-hp- Model 59401A
T
Analyzer
Oscilloscope
Bandwidth:DC to 100MHz
-hp- Model 1740A
T
Signature Analyzer
-hp- Model 5004A
T
* P = Performance
C = Calibration
T = Troubleshooting
O = Operator's Check
a. Output from 30mV to 300V dc.
b. Within .005% full scale accuracy.
c. Short term stability better than .0002% per hour.
4-10. AC Volts Test and Calibration
4-11. For the AC Volts Test and Calibration (AC Volts
Function) use the Fluke Model 5200A/5215A AC
Calibrator. If the calibrator is not available, use a
substitute calibrator that meets the requirements listed
following this paragraph. It is recommended to acquire
the recommended calibrator with the HP-IB (IEEE 488)
option 05 for future computerized Performance Tests
and Calibrations.
a. Frequency Response: 20Hz to 100KHz (and
300KHz at 30V).
b. AC Volts Range: 30mV to 300V ac.
c. AC Volts Accuracy: ± .1%.
4-12. Ohms Test and Calibration
4-13. The 3478A's Ohms Function can be tested and
calibrated in two different ways, at full scale or 1/3 scale.
If the 3478A is to be tested and calibrated at full scale,
use three each of the recommended standard resistors
in series, to form 300, 3000, etc. If the 3478A is to be
tested and calibrated at 1/3 scale, use only one each of
the recommended standard resistors.
4-14. If the recommended resistors are not available,
use substitutes that meet the critical requirements. If a
substitute is not available, you may be able to use a
4-2
Содержание 3478A
Страница 2: ...TM 11 6625 3071 14 A ...
Страница 4: ...TM 11 6625 3071 14 C D BLANK ...
Страница 12: ...TM 11 6625 3071 14 Table 1 1 Specification 1 2 ...
Страница 13: ...TM 11 6625 3071 14 Table 1 1 Specifications Cont 1 3 ...
Страница 14: ...TM 11 6625 3071 14 Table 1 1 Specifications Cont 1 4 ...
Страница 53: ...TM 11 6625 3071 14 1 ...
Страница 54: ...TM 11 6625 3071 14 2 ...
Страница 55: ...TM 11 6625 3071 14 3 ...
Страница 56: ...TM 11 6625 3071 14 4 ...
Страница 87: ...TM 11 6625 3071 14 3478A Figure 7 D 3 Flowchart B 7 D 5 ...
Страница 88: ...TM 11 6625 3071 14 3478A Figure 7 D 4 Flowchart C 7 D 6 ...
Страница 91: ...TM 11 6625 3071 14 3478A Figure 7 D 6 Flowchart D 7 D 9 ...
Страница 98: ...TM 11 6625 3071 14 3478A Figure 7 F 2 Simplified Schematic Of The Input Switching Circuitry 7 F 4 ...
Страница 111: ...TM 11 6625 3071 14 Figure 7 F 17 3478A Simplified Reference Circuitry 7 F 16 ...
Страница 122: ...TM 11 6625 3071 14 Table A 2 A 2 HP IB Worksheet A 4 ...
Страница 137: ...TM 11 6625 3071 14 Figure 7 D 2 Flow chart A 7 D 3 ...
Страница 139: ...TM 11 6625 3071 14 Figure 7 G 2 3478A Block Diagram 7 G 3 ...
Страница 140: ...TM 11 6625 3071 14 Component Locator for Input Circuitry and Ohms Current Source 7 G 4 ...
Страница 141: ...TM 11 6625 3071 14 Figure 7 G 3 Input Circuitry and Ohms Current Source 7 G 5 ...
Страница 142: ...TM 11 6625 3071 14 F G 6 ...
Страница 143: ...TM 11 6625 3071 14 2 Figure 7 G 4 AC to DC Converter 7 G 7 ...
Страница 144: ...TM 11 6625 3071 14 Component Locator for A D Converter and Control Logic 7 G 8 ...
Страница 145: ...TM 11 6625 3071 14 3 Figure 7 G 5 A D Converter and Control Logic 7 G 9 ...
Страница 146: ...TM 11 6625 3071 14 7 G 10 ...
Страница 147: ...TM 11 6625 3071 14 4 Figure 7 G 6 Power Supplies 7 G 11 7 G 12 blank ...
Страница 148: ......
Страница 149: ...PIN NO 057444 ...