TM 11-6625-3071-14
3478A
Table 3-6. 3478A Program Codes
Type
Program
Description
Code
FUNCTION
F1
DC Volts
F2
AC Volts
F3
2-Wire Ohms
F4
4-Wire Ohms
F5
DC Current
F6
AC Current
F7
Extended Ohms (see paragraph 3-118).
RANGE
R-3
most sensitive range on any function
R-2
30mV DC, or most sensitive range on any other
function
R-1
300mV DC, 3OOmV AC, 300OmA DC, 300mA AC,
or most sensitive range on any other function
R0
3V DC, 3V AC, 3A DC, 3A AC, or most sensitive
range on any other function
R1
30V DC, 30V AC, 30 Ohm, or most sensitive range
on any other function
R2
300V DC, 300V AC, 300 Ohm, or least sensitive
range on any other function
R3
3K Ohm or least sensitive range on any other
function
R4
30K Ohm or least sensitive range on any other
function
R5
300K Ohm or least sensitive range on any other
function
R6
3M Ohm or least sensitive range on any other
function
R7
30M Ohm or least sensitive range on any other
function.
RA
Autorange
DIGITS
N3
3 1/2 Digits Displayed
DISPLAYED
N4
4 1/2 Digits Displayed
N5
5 112 Digits Displayed
TRIGGER
T1
Internal Trigger
T2
External Trigger
T3
Single Trigger
T4
Hold Trigger
T5
Fast Trigger (see paragraph 3-116)
AUTOZERO
Z0
Autozero Disabled
Z1
Autozero Enabled
DISPLAY
D1
Normal Display Operation
D2
Display Text and update Annunciators
D3
Display Text and turn off Annunciators
HOME
HO
Equivalent to "F1T4R-2RAZ1N4"
COMMAND
H1
Equivalent to "F1R-2RAZ1N4T3"
H2
Equivalent to "F2R-2RAZ1N4T3"
H3
Equivalent to "F3R-2RAZ1N4T3"
H4
Equivalent to "F4R-2RAZ1N4T3"
H5
Equivalent to "F5R-2RAZ1N4T3"
H6
Equivalent to "F6R-2RAZ1N4T3"
H7
Equivalent to "F7R-2RAZ1N4T3"
MISC
B
Read Status of the 3478A in Binary
COMMANDS
See paragraph 3-119.
C
Calibrate (see Section IV of this manual)
E
Read the Error Register. See paragraph 3-1 20.
K
Clears Status Register
M
Set SRO Mask (Mdd, where dd is the octal code of
the bits)
S
Front/Rear Switch Position
(0 = Rear, 1 = Front)
77) can be sent to program the mask. The mask can be
cleared by sending: "MOO". Remember, the mask is
only set to output a Require Service Message (not the
Status Byte) for a certain SRQ condition.
3-105. Clearing Status Register (Status Byte). Bits 2
through 5 and 7 of the register can be cleared by sending
program code "K", and by a device CLEAR message. Bit
0 (Data Ready) is not cleared since it represents the
3478A’s current status (the bit is cleared when the 3478A
outputs its reading). Bit 6 is the SRQ bit which can be
cleared by a Serial Poll or if the SRQ Mask is set to 00.
3-106. Power-On SRQ. This feature is enabled by
setting the Power-On SRQ switch to On (switch #3 on
the rear panel in the up position). When the feature is
enabled, the 3478A Requires Service (generates an
SRQ) each time power is cycled or a reset condition is
generated by the instrument (e.g. due to an instrument
failure or selecting the Self-Test).
3-107. Sending Data to the Display. Up to 12 ASCII
characters can be displayed at a time by the 3478A’s
display. The legal characters are decimal 32 through 95
of the 128 ASCII characters. Only upper case letters
and numbers can be displayed. Lower case letters
generate characters which do not resemble the letters.
Commas, periods, and semicolons can go between
characters. The ASCII characters can be sent in two
different modes. The following paragraphs explain the
two modes.
3-108. In one display mode, the display is continuously
updated. This is enabled by program code "D2" (e.g.
"D23478A DMM" where "3478A DMM" is the message
displayed). In this mode, the annunciators continue to be
updated. If more than 12 characters are sent to the
display, the extra characters are ignored until a control
character is received. If the control character is different
than HT (Horizontal Tab), VT (Vertical Tab), LF (Line
Field), CR (Carriage Return), or FF (Form Feed), a
Syntax Error is generated. A "D2" code locks the display
until a "DI" (Selects Normal Display) is sent, a CLEAR
message is sent, an error condition occurs, or a front
panel button is pressed.
3-109. The other display mode is selected by sending
code "D3". This mode is the same as the previous
mode, except the display is not updated and all
annunciators are turned off. The text on the display
remains on for about 10 minutes and then the display
goes blank. The display can be restored by sending any
of the display codes (DI, D2, or D3) or by pressing the
LOCAL button (if not disabled by the LOCAL LOCKOUT
Message).
3-110. Home Commands. The Home Commands
(when sent to the 3478A over the HP-IB) are used to set
the 3478A into a predefined operating state. The
program code is "Hn", where n is the number which,
3-12
Содержание 3478A
Страница 2: ...TM 11 6625 3071 14 A ...
Страница 4: ...TM 11 6625 3071 14 C D BLANK ...
Страница 12: ...TM 11 6625 3071 14 Table 1 1 Specification 1 2 ...
Страница 13: ...TM 11 6625 3071 14 Table 1 1 Specifications Cont 1 3 ...
Страница 14: ...TM 11 6625 3071 14 Table 1 1 Specifications Cont 1 4 ...
Страница 53: ...TM 11 6625 3071 14 1 ...
Страница 54: ...TM 11 6625 3071 14 2 ...
Страница 55: ...TM 11 6625 3071 14 3 ...
Страница 56: ...TM 11 6625 3071 14 4 ...
Страница 87: ...TM 11 6625 3071 14 3478A Figure 7 D 3 Flowchart B 7 D 5 ...
Страница 88: ...TM 11 6625 3071 14 3478A Figure 7 D 4 Flowchart C 7 D 6 ...
Страница 91: ...TM 11 6625 3071 14 3478A Figure 7 D 6 Flowchart D 7 D 9 ...
Страница 98: ...TM 11 6625 3071 14 3478A Figure 7 F 2 Simplified Schematic Of The Input Switching Circuitry 7 F 4 ...
Страница 111: ...TM 11 6625 3071 14 Figure 7 F 17 3478A Simplified Reference Circuitry 7 F 16 ...
Страница 122: ...TM 11 6625 3071 14 Table A 2 A 2 HP IB Worksheet A 4 ...
Страница 137: ...TM 11 6625 3071 14 Figure 7 D 2 Flow chart A 7 D 3 ...
Страница 139: ...TM 11 6625 3071 14 Figure 7 G 2 3478A Block Diagram 7 G 3 ...
Страница 140: ...TM 11 6625 3071 14 Component Locator for Input Circuitry and Ohms Current Source 7 G 4 ...
Страница 141: ...TM 11 6625 3071 14 Figure 7 G 3 Input Circuitry and Ohms Current Source 7 G 5 ...
Страница 142: ...TM 11 6625 3071 14 F G 6 ...
Страница 143: ...TM 11 6625 3071 14 2 Figure 7 G 4 AC to DC Converter 7 G 7 ...
Страница 144: ...TM 11 6625 3071 14 Component Locator for A D Converter and Control Logic 7 G 8 ...
Страница 145: ...TM 11 6625 3071 14 3 Figure 7 G 5 A D Converter and Control Logic 7 G 9 ...
Страница 146: ...TM 11 6625 3071 14 7 G 10 ...
Страница 147: ...TM 11 6625 3071 14 4 Figure 7 G 6 Power Supplies 7 G 11 7 G 12 blank ...
Страница 148: ......
Страница 149: ...PIN NO 057444 ...