Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17-3 Query/Retrieve Verification (cont’d)
Follow the Steps Below to Test the DICOM Query/Retrieve Workflow (cont’d)
-
If the test fails, a pop-up dialog reporting the results as “0 of 1 OK” is displayed. In the example
the warning message pop up almost at once, and it states that for the Server
Check (Ping), 1 of 1 is OK but the Device Check failed (0 of 1 OK). This indicates that the
network connection functions OK, but the application on the remote server is not answering.
If the test fails directly, as described in this example, the AE Title is probably wrong.
A red “X” to the left of the Selected devices indicates that the test failed, but it doesn’t indicate the cause.
-
If the Check button fails after a long time (corresponding to the timeout), the Warning will look
like the example in
, where both tests failed.
In this case, the IP address or Port Number is probably wrong.
If Retrieve is Failing ...
If Retrieve is failing but Query is functioning, verify if the scanner has been registered as an receiver on
the server.
Figure 3-86 Test (Check) Fails directly
Figure 3-87 Test (Check) Passed
Figure 3-88 Both Tests Failed
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......