GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
3 - 106
Section 3-10 - Connectivity Setup - Software v3.x
3-10-10-3
Query/Retrieve Verification
Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection
1.) Select the “Smiley” button to Ping the server.
NOTE:
The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM
ping), so it does not check AE title or port number.
2.) If the network connection to the server is OK, it will be illustrated by a smiling “Smiley”
.
A sad Smiley indicates that the network connection is failing.
Typical causes:
-
Network cable not connected.
-
Configuration error(s).
3.) When ready, select OK to close the DICOM Query/Retrieve properties and save changes.
Figure 3-112 DICOM Query/Retrieve properties
SMILEY
OK