GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
8 - 40
Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-6
Backup - Software Version v3.x
The Backup function enables the user to:
•
Make backup of the patient archive (patient data and reports).
•
Move images from the local hard drive to MO disks.
•
Make backup of the system configuration.
8-8-6-1
Move Images From the Local Hard Drive to MO disks.
The
Move Images
function is used to move archived images from one volume/media to another and the
deletion of the moved images from the original volume. The pointers to the database are restored in the
new volume/media.
NOTE:
This function copies image data only, no database information is copied.
NOTE:
When moving images, be sure to have enough formatted MO disks available.
1.) Format the MO disks you’re going to use for the backup, see
4-2-7 "Formatting Removable Media
- Software Version v3.x and above" on page 4-17
.
2.) Insert a dedicated and formatted MO disk into the MO drive.
3.) Press
CONFIG
(
F2
) and log on as
adm
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
.
4.) Select the Admin setup category.
5.) Select
Move oldest images
to move images from the local hard drive onto MO disks.
6.) Select the MO disk as destination.
CAUTION
Use blank dedicated MO disks when doing backup. The backup will OVERWRITE the existing
data on the MO disk.
Figure 8-35 The Admin screen
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......