Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-12
Remote Path
Remote Path is used for Save As, Export from Q-Analysis, and for exporting error logs with the <Alt+D>
command.
Follow this procedure to Configure Vivid 7 to use the
Remote Path
.
1.) Press
Config
(
F2
) and log on as
ADM
, see
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) If not already selected, select
Connectivity
from the bottom row of “buttons” on the screen.
3.) Select the
Tools
tab.
4.) Enter the
Remote Path
.
Example:
\\3.187.187.30\EXPORT\804
5.) Use the
Check
button to verify the connection.
Figure 3-51 Tools screen
REMOTE PATH
CONFIGURABLE
REMOTE PATH USER