3 - 78
Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17
Query/Retrieve (Q/R) Setup for Software v7.x/v6.x/v5.x/v4.x
3-9-17-1
Overview
The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM
server for further analysis on the Vivid 7.
NOTE:
You may have to set up Vivid 7 as a destination on the server.
3-9-17-2
Query/Retrieve Setup on the Vivid 7 with Software
v7.x/v6.x/v5.x/v4.x
1.) Press
CONFIG
(
F2
) and log on as
adm
, see
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) Select Connectivity
3.) Select Dataflow
4.) Select Query Retrieve from the Name pull down menu
Figure 3-77 Select Query Retrieve
NAME PULLDOWN
MENU
QUERY RETRIEVE