Chapter 3 - System Setup
3 - 77
3-9-16-3
Create a New Dataflow with Worklist and DICOM Storage (for CA-1000 sites)
This dataflow is intended for Vivid 7 scanners in a DICOM network, where the data should not be stored
locally on the scanner. A site with the integrated EchoPAC PC and CA-1000 is an example of this.
To create a new dataflow with Worklist and DICOM Storage, do the following steps:
1.) Add a new dataflow in Config screen, Connectivity, Dataflow tab. Enter a name, e.g.
Worklist -
DICOM Storage
.
2.) Select
OK
.
3.) Add the Worklist input. Configure (at least)
IP address
,
AE title
and
Port Number
.
4.) Select
OK
.
5.) Add the Dicom Storage output.
6.) Configure (at least)
IP address
,
AE title
and
Port Number
.
If the site has integrated EchoPAC PC
and CA-1000, enable
Allow SR
and
Allow SR Private Data
.
7.) Select
OK
.
8.) If this dataflow should be the default at start-up, enable
Default
for the dataflow.
Figure 3-76 Dicom SR Settings
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......