Section 10-5 - System Maintenance
10-5-2-2
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
10-5-3
Input Power
10-5-3-1
Mains Cable Inspection
Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks
Step
Item
Description
1
DVR/VCR
Verify record/playback capabilities of the unit: Clean heads (VCR ONLY) and covers if necessary.
2
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3
Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
5
InSite/iLinq
Verify that InSite is functioning properly. Ensure two-way remote communications.
(Warranty & Contract Customers only)
6
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
7
ECG
Verify basic operation with customer
Table 10-7 Mains Cable Inspection
Step
Item
Description
1
Unplug Cord Disconnect the mains cable from the wall and system.
2
Inspect
Inspect it and its connectors for damage of any kinds.
3
Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no
strands may cause a short circuit.
4
Verify
Inlet connector retainer is functional.
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......