GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
1 - 52
Section 1-5 - Labels Locations
1-5-19
Label, Internal Connections (Int.Conn.)
Located on the outside of the Front-End Card Rack’s cover (inside unit). Different versions of the label
have been used since production start.
Figure 1-7 Label on FEP2 (w/RFI)
K1
K2
K3
Cable and Connector numbering and naming system used:
IIO
= Internal IO module
EIO
= External IO module
BEP
= Backend Processor
FEP
= Frontend Processor
ACP
= AC controller module
ACD
= AC distribution box
ACT
= AC isolation transformer module
BEPIO
= IO module on the BEP
Axx :
connectors on IIO with other connections than the BEP
Bxx :
connectors on IIO that are connected to BEP
Cxx :
connectors on BEP, except for the BEPIO
Dxx :
connectors on BEPI O
Exx :
connector on the BEP power s u p p l y
Fxx :
connectors on AC P
Gxx :
connectors on FEP
Hxx :
external accesable connectors (Doppler probe)
Kxx :
connectors on MODEM
Mxx :
connectors on ACT
Nxx :
connectors on ACD
Lxx :
connectors on EI O
External IO module
6
9
1
5
M
14
6
9
1
5
M
Analogue
phone
Connect
to InSite
MODEM
RS-232
Connect
to InSite
MODEM
Service
use ONLY
(remote
power on/off
control)
L1
L2
L3
BF-6
4
RF
I
DC-pow
er
supply
TX-pow
e
r
supp
ly
Board location in Front End Processor (FEP)
Monito
r
Periphera
ls
Mode
m
Back End
Processo
r
Front End
Ra
ck
115V
115V
or
230V
230V
230V
230V
From Isolation
transformer
230V
115V
WARNING !
Live Voltage inside,
do not open.
AC power control modu l e
To Isolation
transformer
115V or 230V inlet
voltage selection
Voltage Selec tion
on Peripherals
115V = 100-1
20V
230V = 220-2
40V
(½ or 2 times i
nlet
voltag
e)
H 1
G1
F1
F2
F3
F4
F5
F6
F7
F8
F9
Isolation
transformer
modul e
M1
69
15
M
230V
115V
Fa
n
contr
ol
81
15
9
F
A13
81
15
9
F
A12
11
3
14
25
M
A16
18
91
5
M
A15
69
15
M
A14
A12
Spare power (future option)
A13
Power to 3D box (future option)
A14
AC power control
A15
Signals to rotation adapter (future option)
A16
Signals to 3D box (future option)
connector
to EIO
1
25
A
F
P30
Connecto
r t
o
motherboa
rd
P31
P32
IIO seen from F
EP s
ide
Part no. : FC314311
Rev. : 01
60,0
Cable to
C15
Modem
MT5634ZBA
Modem
power
suppl y
K4
G4
G3
AC
Distribution
bo x
N1
N2
N3
N4
N5
RX-1
28
Relay board
M
F
L4
L5
L6
L7
L8
L9
L10
L11
L12
L13
L14
L15
L16
L17
External IO module
(Front v iew)
L4:
Audio out (right)
L5:
Audio out (lef t)
L6:
Trig out
L7:
Foot swit ch
L8:
Serial port
L9:
Remote expose #1
L10:
Remote expose #2
L11:
Analog modem
L12:
Composite video output
L13:
B&W video output
L14:
S-Video output
L15:
SVGA out put
L16:
USB
L17:
Ethernet
RESET
TX-128
BF-6
4
G2
TX128 (optional)
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......