Chapter 7 - Diagnostics/Troubleshooting
7 - 73
7-8-3-10
Image Port Test
When selecting the ImagePort Test, the scanner issues a test command to the local software running
on the I960RP processor on the Image Port board.
The test command will start the following tests;
•
test the interrupt based command/replay mechanism
•
reset the ImagePort board
•
test the PipeLink data fifo
•
test the video data fifos
•
test the PipeLink control fifo
•
test the video control fifo
•
test the video grabber chip
Results;
•
Green light: - OK
•
Red light: - Failed
7-8-3-11
Chain (Signal Chain, RLB/RX/BF Test)
NOTE:
A Test Attenuator, Part Number AA200215 is needed for this test.
Remember to restart the Vivid 7 after the diagnostics session.
,
Table 7-11 Signal Chain (TX/RLB/RX/BF) Test
Test (Key) Sub-tests
Description
Chain
Signal Chain Test (Chain)
The Signal Chain/ Data Flow test checks for errors in the TX, RLY, RX and BF chain of data
flow.
Test All
Not Used.
TX/RLB/RX/
BF
Tests data flow in Front-End Transmit Receive signal chain.
If an error occurs with this test, testing should be continued in the following order;
1. BF digital and analog test
2. RX test
3. RLY test.
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......