Chapter 4 - Functional Checks
4 - 19
4-2-10
Archiving and Loading Presets - Software Version v2.x/v3.x
NOTE:
Always save presets before any software reload. This action is necessary to ensure that the
presets loaded after the software reload are as up–to–date as possible.
4-2-10-1
Archiving Presets to an MO Disk
NOTICE
Presets shouldn’t be saved on the same MO disk as images, because the Archive Menu lists the
images but not the presets stored on a MO disk.
Table 4-3
Saving Presets
STEP
TASK
FUNCTION
1.
Insert an empty, formatted MO disk
labelled “PRESETS” and marked with
Vivid 7’s serial number and the date,
into the MO Drive.
For more information about
formatting, see
Removable Media - Software Version
v3.x and above" on page 4-17
2.
Go to Config (F2) -> Admin -> Backup
The Backup window is displayed.
3.
Select the Check box “User defined
Configuration”.
4.
Select the MOD as destination.
5.
Press
BACKUP NOW
.
A copy of the Presets is saved on the MO disk.
6.
Press
ALT
+
E
on the alphanumeric
keyboard.
The MO disk is ejected.
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......