GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 5 - Components and Functions (Theory)
5 - 117
Section 5-8
External I/O
5-8-1
General Description, External I/O
The External I/O is the interface between the scanner and all external devices.
Several signal are fed from the Internal I/O module to the External I/O module. The External I/O module
contains data buffers and a galvanic isolation section.
Different types of signals are fed to connectors accessible on the External I/O:
•
R, G, B and Sync are provided for external RGB monitors or printers.
•
Composite video, black and white video and S-VHS video are buffered and fed to output connectors
for external monitors and printers.
•
Two serial RS-232 ports are provided for external communication with e.g. the System CPU. One
of the RS-232C ports are located inside the scanner, and was intended for use by a modem, the
other is located on the EIO rear panel. These serial ports have never been used on Vivid 7.
•
An Ethernet signal from the BEP is fed to an Ethernet connector. It is used for connectivity, including
DICOM Network.
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......