Chapter 5 - Components and Functions (Theory)
5 - 131
5-11-2-3
Distribution of Voltages via AC Controller
The 20 VAC is routed via the fan regulator to the system fan.
Both 115 VAC and 230 VAC are routed via fuses.
•
115 VAC is routed to the Monitor outlet (MONITOR)
•
An AC Voltage selector is used to select the correct voltage for the Peripherals. The selected
voltage (230 VAC or 115 VAC) is routed to the Peripherals AC outlet (PERIPH’S)
•
230 VAC is distributed to;
-
Modem AC outlet (MODEM)
-
Back-End Processor AC outlet (PC)
-
Front End Card Cage AC outlet (RACK)
Figure 5-84 Connector Pin-out - AC Connectors on AC Controller