GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4 - 43
4-3-8
Color Mode Checks
4-3-8-1
Introduction
Color Flow screens is 2D or M Mode screens with colors representing blood or tissue movement.
Color Flow may be selected both from 2D mode or from M mode or a combination of these.
4-3-8-2
Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-
"Probe Connection" on page 3-21
for info about connecting the probes.
-
2.) Turn ON the scanner.
-
on page
for info about booting the scanner
The 2D Mode window is displayed (default mode).
4-3-8-3
Color 2D Mode
Press
COLOR
.
4-3-8-4
Color M Mode
1.) Press
MM
(M Mode) to bring up an M Mode display on the screen.
2.) Press
COLOR
.