Chapter 8 - Replacement Procedures
8 - 37
Restore Patient Archive/System Config. - Software v7.x/v6.x/v5.x/v4.x (cont’d)
8.) Select
Restore Now
. Depending on the selection, one or two
Restore confirmation
windows are
displayed:
9.) Ensure that the correct source is selected an select
OK
. The selected items are copied to the
systems. If items from the
System Configuration
are restored the system needs to be rebooted.
The Reboot system window is displayed.
10.)Select
OK
to reboot the system.
8-7-15
Verification - Functional Checks
•
Please perform the tests specified in
Section 8-20 "Verification - Functional Checks" on page 8-105
Figure 8-33 The Restore Confirmation window
Figure 8-34 The Reboot System window