Chapter 7 - Diagnostics/Troubleshooting
7 - 29
7-6-5-4
Control Frame
Contains the user interface elements used for:
•
Diagnostic control
•
Operator feedback
a.) Execute Button
This button has two modes - each with appropriate text:
*
Execute - to start the diagnostic, and
*
Abort - to stop a diagnostic
b.) Loop Count
This is an editable text field that will only accept numeric values with 4 digits or less. When the
button described in
step
(above) is configured as an “execute” button and
pressed, the loop count field will be queried to determine the number of times to execute the
diagnostic.
c.) Progress Indicator
Displays a graphical progress indicator for the user.
d.) Short Text Message
Displays either a starting message or aborting message, as well as the diagnostic completion
status.
e.) Control Frame Background Color
Initially gray, the Control Frame background color changes upon completion of a diagnostic
to indicate completion status.
*
Fail = Red
*
Pass = Green
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......