Chapter 8 - Replacement Procedures
8 - 35
8-7-13-3
Verify TCP/IP Settings - Software Version v7.x/v6.x/v5.x/v4.x
1.) From the top of the Connectivity screen, select the
Tcpip
tab.
2.) Verify that all settings on the TCPIP tab matches those recorded in
Table 8-6 "Record settings from
Tcpip screen - software version v7.x/v6.x/v5.x/v4.x" on page 8-21
3.) Select
Save settings
.
4.) Select
Ok
to continue.
8-7-13-4
DC Offset Calibration
1.) Disconnect all probes.
2.) From the upper section of the window, select
Test
.
3.) Then select
Go to System Test.
4.) From the Module selection block, select
DC Offs. Cal
.
5.) Reboot the System when the DC Offset Calibration has been completed.
8-7-13-5
Verify System Software Version
1.) Press
CONFIG
(
F2
) and log on as
adm
, see
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) Select
About
tab.
3.) Verify that the software versions are the same as what were recorded in
.
Figure 8-31 TCP/IP Example Settings
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......