Chapter 4 - Functional Checks
4 - 27
Moving and Transporting the Vivid 7 (cont’d)
•
Place the probes in their carrying case.
•
Lock Operator Panel (Control Console) in place.
The Operator Panel (Control Console) must be locked into the center (x/y) position as indicated on
the label on the Front Handle, before transporting the system.
•
Eject any Magneto Optical disk, CD or DVD from their drive.
•
Secure the unit in an upright position.
•
Lock the casters (wheels) (brake)
•
DO NOT use the Operator Panel (Control Console) as an anchor point.
NOTE:
Special care should be taken when transporting the unit in a vehicle:
Please refer to the Vivid 7 Unpacking/Packing Procedure, Direction Number FC050172 - Section 4-3
“Loading and Transporting the Unit in a Van”, if you need to transport the Vivid 7 by vehicle, and don’t
have the transportation cage available.
Figure 4-24 Label on Front Handle
OR
OR