7 - 6
Section 7-5 - Gathering Troubleshooting Data
7-5-3
Collect a Trouble Image with Logs
If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture
of the monitor, system presets and several log files in a date and time stamped zip file.
NOTE:
Power Supply and Temperature logs are not currently being updated by the Vivid 7.
This Alt-D function is available at all times.
When Alt-D is pressed, a menu box appears that allows for;
•
a place to enter a description of the issue
•
a check box to indicate a System lockup
•
a choice to Export to a pre-formatted MOD, CD-RW or save to the Export directory D: drive (for
remote viewing through InSite).
NOTE:
NOTE: You
MUST
select one of the available devices as the destination device if it is to be
different than the default Export directory on the hard drive.
The screen capture is a bitmap which eliminates the possibility of artifacts from compression.
Figure 7-2 ALT-D Dialog Box
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......