GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 3 - System Setup
3 - 67
3-9-14-6 DICOM Storage Properties Adjustments (cont’d)
DICOM SR
If SR is enabled, all M&A that is supported by DICOM is sent to the DICOM server (in a separate job).
Allow SR Private Data (introduced with software versions v6.0, v5.2 and v4.2.)
By default, this function is disabled.
Allow SR Private Data
is intended for receivers that can handle “Private Data”, like the CA-1000.
In the examples in Figure 3-67 and Figure 3-68 (below),
Allow SR Private Data
has been enabled.
Figure 3-67 Dicom Storage properties
Figure 3-68 Dicom Media properties