4
Select whether or not to back up the setup data at discharge.
Selecting ON will back up the alarm settings and
display configuration settings when the discharge
process is performed.
Selecting OFF will initialize the alarm settings and
display configuration settings.
5
Select the time interval for “NIBP Auto Mode at Power ON/Discharge”.
Pressing the key will display the NIBP auto mode time
interval selection window.
Select the time interval.
6
Set the “Label Setup at Discharge/Change of Disp. Mode”.
Backup : At discharge and at change of display mode,
current BP label, TEMP label setting will be applied
regardless of the label setting made on the display
mode setup menu.
Disp. Mode : At discharge and at change of display
mode, BP label, TEMP label, color setting made on the
display mode setup menu will be applied.