6
-
33
●
Display Format of BP Numeric Data
The display format of BP numeric data can be selected from systolic / diastolic / mean, systolic /
diastolic, or mean BP. If the BP label is CVP, PAP, ICP, or IAP, this selection is not possible.
1
Select the display format.
S/D/M will display the systolic / diastolic / mean pressure.
S/D will display the systolic / diastolic pressure.
M will display only the mean pressure.
C A U T I O N
The undisplayed BP data (S/D/M) will not generate a BP alarm or be
displayed in the table data. Select the appropriate display format according to
the monitoring purpose.
●
Mean BP Waveform Display
The mean BP waveform can be set to be continuously displayed on the home display.
1
Select ON or OFF for mean BP waveform display.
Selecting ON will display the mean BP waveform and
“Mean Wave” will be displayed inside the numeric data box.
6
Inva
sive Bloo
d Pressu
re
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...