C A U T I O N
y
The accuracy of MGU-801P/MGU-802/MGU-803 may be affected at
extreme temperatures. Do not store them at extreme temperature.
Temperatures exceeding specified storage temperatures (-10 to 60
°
C)
could damage the units.
y
After the MGU-801P/MGU-802/MGU-803 is stored in low temperature,
when turning on the device, it may require additional warm-up time.
z
Multigas Monitoring (PHASEIN
®
IRMA Probe)
y
Never sterilize or immerse the IRMA probe in liquid.
y
The IRMA oxygen sensor cell and IRMA airway adapters are non-sterile
devices. Do not autoclave the devices as this will damage them.
y
Do not apply tension to the probe cable.
y
Do not operate the IRMA probe outside the specified operating
temperature environment.
y
Do not leave depleted oxygen sensors mounted in the IRMA probe,
even if the probe is not in use.
y
The O
2
sensor needs to be replaced every 4 months. Replace within 4
months after opening the package even if it is not used. When the
“Replace O
2
sensor” message is displayed, O
2
measurement cannot be
performed. Make sure to use the sensor before the expiration date
indicated on the package.
y
The disposable airway adapter should be opened just before use. Do
not sterilize it.
y
Do not reuse the disposable airway adapter.
y
The airway adapter should be attached with the thicker side facing to the
patient. If attached oppositely, it may damage the IRMA probe or airway
adapter.
y
If the O
2
concentration is low (in such case if the airway adapter is
removed near the nose/mouse where it is affected by expired gas), O
2
sensor calibration will not be properly performed. After removing the
airway adapter, locate it away where it will not be affected by expired
gas.
z
Alarm
y
Pay attention not to set the alarm volume too low to avoid missing any
important alarms.
y
Alarm messages will be displayed according to the priority. (Level 1
→
Level 2
→
Level 3
→
Level 4)
y
For the same alarm level, the alarm message for the newer alarm will be
displayed.
y
On the DS-7000, HR alarm and PR alarm cannot be set to ON at the
same time.
y
Whether to use the SEC alarm function and its threshold selection
should be based on the patient’s clinical indication portent and medical
evaluation.
y
If the SpO
2
alarm and SEC alarm setup is set to OFF, the SEC alarm
integral value will be set to 0.
z
System Configuration
y
If the time/date is not correctly set, or if changed during monitoring,
malfunction may occur to NIBP measurement, periodic recording, trend,
NIBP list data, and age calculation from the birth date.
y
When the waveform and numeric data display for each parameter is set
to OFF, the alarm generation and table/trend input for the corresponded
parameter will be also set to OFF.
y
If the display of waveform / numeric data labeled as BP1 or ART is set to
OFF, the pulse rate derived from BP will not be displayed either.
y
When the waveform and numeric data display for SpO
2
is set to OFF, the
pulse rate derived from SpO
2
will not be displayed either.
y
When the waveform and numeric data display for CO
2
is set to OFF, the
respiration rate derived from CO
2
will not be displayed either.
xxiii
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...