4
-
77
Error Message
CF card slot : OUT
Cause : CF card is not inserted or not correctly set in the CF card slot.
Solution : Set the CF card into the CF card slot.
Invalid CF card.
Cause : Unspecified CF card is used.
Solution : Set a specified CF card into the CF card slot.
No data on the CF card.
Cause : There is no data on the CF card.
Solution : Check if the correct CF card is being used, or rewrite the data on the CF card.
CF card error.
Cause 1 : An error has been detected when writing/reading data on the CF card.
Solution : If the error has been detected during writing, try again.
If the error has been detected during reading, data might not be correctly written on the CF
card. Rewrite the data after formatting and try the procedure again.
Cause 2 : The software version of the DS-7000 main unit is older than that of the data stored in the CF
card.
Solution : The data from a newer software version cannot be read.
Update the software version of the DS-7000.
Cause 3 : There is no more space on the CF card to write the data.
Solution : Format the CF card.