Message List
Message
Description
Status Message
Wait
Preparing for measurement.
Also displayed when the catheter relay cable is not connected to the CO
module, or when the thermodilution catheter is not connected.
Ready
Ready to begin the measurement.
Busy
In process of measurement.
Finished
Measurement is finished.
Result Status
CO OK
CO is correctly measured.
Upper Fault
Measurement error
・
Blood temperature is out of the measurable range after the injection.
・
Thermistor connector or relay cable is not properly connected.
・
Line is cut on the sensor or relay cable.
Peak Fault
Measurement error
・
Peak of the thermodilution curve can not be detected.
・
Thermistor connector or relay cable is not properly connected.
・
Line is cut on the sensor or relay cable.
Lower Fault
Measurement error
・
Blood temperature has not returned to a stable condition after
measurement.
・
Thermistor connector or relay cable is not properly connected.
・
Line is cut on the sensor or relay cable.
Sensor Err.
Measurement error
・
Thermistor connector and relay cable is not properly connected.
・
Line is cut on the sensor or relay cable.
Over Range
Measurement error
・
CO value is out of measurable range.
The result status will be displayed for 30 seconds after completion of measurement.
7
Cardia
c Outp
ut Measu
rem
ent
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...