Electrosurgery Safety
W A R N I N G
The monitoring system contains protection against interference generated by
electrosurgical instruments. However, operating conditions, surgery site with
respect to the location of ECG electrodes, or the type of instrument used, may
cause noise on the ECG. The noise is generated at the tip of an electrical knife
and is difficult to completely eliminate because of the frequency components of
the ECG. To reduce electrosurgical interference, take the following
precautions:
Location
Locate the electrosurgical unit as far as possible from this unit and the
patient cable. This will help reduce interference on the ECG through the
monitor or cables.
Power Supply
Connect the electrosurgical unit to a power supply that is different from that
of the monitor. This will help prevent interference through the power cable.
Electrode Placement
The amount of interference is considerably different depending on the
electrode position and surgery site. Place the ECG electrodes as far away
as possible from the surgery site and the ground plate. Do not place
electrodes in the path between the surgery site and the ground plate. If the
electrodes are placed in this path, the amount of interference will be quite
large. Position (+) and (–) electrodes as close as possible to each other.
Ground Plate
When using electrosurgical instruments, make sure the contact between
the patient and the ground plate is secure. If the contact is poor, the patient
may suffer a burn at the electrode site.
Precautions about Magnetic Resonance Imaging
W A R N I N G
z
Do not operate this equipment in magnetic resonance imaging (MRI)
environments.
z
When conducting MRI test, remove the electrodes and sensors connected
to the patient (test subject).
The local heating caused by the induced electromotive force may cause
burn injury to the patient (subject). For details, refer to the operation
manual for the MRI testing device.
xiv
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...