Patient Classification
The selection of patient classification affects the accuracy of NIBP measurement, HR measurement,
and RR measurement. Also the delay time to generate the measurement data alarm will change
according to the patient classification.
Adult
Child
Neonate
NIBP measurement range
10 to 260mmHg
10 to 180mmHg
10 to 130mmHg
HR
0bpm, 12 to 300bpm
0bpm, 30 to 300bpm
Monitor
0.5 to 40Hz
1.6 to 40Hz
ESIS
1.6 to 15Hz
1.6 to 15Hz
Filter
Mode
Diagnosis
0.05 to 100Hz
1.6 to 100Hz
Impedance Respiration
1.5Hz
2.5Hz
Alarm delay time
5 sec.
0 sec.
The alarm delay time is a function to prevent frequent generation of the measurement data alarm by
holding the alarm generation for the duration of each delay time.
The alarm delay time applies to the measurement data alarm for the following parameters; HR/PR, BP,
RR, SpO
2
, TEMP, TACHY, BRADY.
W A R N I N G
The monitor determines the detection algorithms for QRS and NIBP according
to the selected patient classification. Make sure the proper selection is made.
1
Select the patient classification.
Select Adult , Child , or Neonate .
Admitting a Patient
5
2
The selected patient classification will be displayed on the home display.
Patient Classification
5
-
5
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...