6
-
28
PCWP Measurement
When PAP is set as BP label, the mean value can be displayed as PCWP (Pulmonary Capillary
Wedge Pressure).
1
Set the BP label to PAP, and display the BP menu.
If the BP label is set to PAP, the PCWP key will be
displayed on the BP menu.
2
Press the PCWP key.
The PCWP measurement display will appear.
3
Press the Wave Read key.
PCWP waveform and respiration waveform will be displayed.
The cursor point indicates the current mean pressure.
4
Use the ,
keys to manually set the PCWP value.
Move the cursor to manually set the PCWP value.
Select the waveform scale from 20 or 50mmHg for optimum view.
5
Press the Input key.
Press the Input key when the PCWP value is correctly set.
The PCWP value will be displayed inside the PAP (BP label) numeric data box with the measurement
time.
Also, this value will be input to the trend data.