Alarm Mode Setup
This section explains the procedure to program the alarm mode.
About the Alarm Mode
On the DS-7000, 5 patterns of alarm mode for each patient classification can be programmed
according to the monitoring purpose.
Setting all the alarm condition each time the patient is admitted may be troublesome.
To simplify this procedure, 5 patterns of alarm mode other than the default setting can be programmed
according to the monitoring purpose.
The alarm mode can be programmed for each patient classification (adult, child, neonate) which will
allow to easily change the alarm setting when the patient classification is changed.
By preprogramming the setups to each alarm mode, alarm setups at admitting procedure can be
simplified by just selecting one alarm mode.
It is recommended to program the alarm mode in rough classification such as patient’s age, monitoring
purpose (ICU or surgery), and if necessary, perform a unique setup for each patient.
Alarm Setup
Alarm Setup
Alarm Mode 1
Alarm Mode Setup
Adjust the alarm setup on the alarm setup
menu according to the patient’s condition.
Current alarm setup
Select an alarm mode when admitting a
patient.
Program an alarm mode.
9
Alarm Mod
e Setup
9
-
3
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...