C A U T I O N
z
When connecting to the DS-LAN network, perform “DS-LAN Setup” in the
Initial Settings and restart the system before connecting the LAN cable.
z
If performing wired network transmission, configure the display so that the
numeric data corresponded to the waveform is displayed. If not, the
displayed waveform or numeric data may not be transmitted.
z
The Monitor ID is factory set to 000. If connected to the wired network with
the ID unchanged, monitoring on the central monitor will not be possible.
z
When connected to the wired network, make sure that there are no other
bedside monitors with the same Bed ID/Monitor ID. If there are more than
one bedside monitors with the same Bed ID/Monitor ID, the duplicated
bedside monitors cannot be monitored on the central monitor.
z
Make sure to set the Monitor ID in the following range.
y
For DS-LAN
II
network: 001 to 048
y
For DS-LAN
III
network: 001 to 100
z
On a wired network (DS-LAN
II
/
III
) system, if the discharge procedure is
performed on the central monitor, the alarm setting on the bedside monitor
will be initialized to the value set on the “Admit Setup” of the central monitor.
z
As the DS-7000 does not have the arrhythmia template display and 12-lead
ST display function, these displays on the central monitor will not be
corresponded.
z
If connected to a wired network, time/date will be the same with the central
monitor. Even if the time/date is changed on the DS-7000 system, it will be
corrected to the time/date of the central monitor.
z
On a wired network, the alarm generated on the bedside monitor will be
output to the network with a maximum delay of 1 second, and to the central
monitor with a total delay of 2.5 seconds.
z
If the Alarm Silence key is pressed during generation of upper and lower
NIBP alarm, the alarm sound and message will be suspended on the central
monitor connected to the wired network.
z
In case of DS-LANII network, if the “HR/PR Alarm Source” is PR, and “PR
Source” is BP, the ECG waveform will not be transmitted. On the central
monitor, PR-IBP value will be displayed for HR. However, the HR value from
ECG will be displayed on the NIBP list and ST measurement table. Refer
also to the operation manual for the respective central monitor.
In case of DS-LAN
III
network, refer to the operation manual for the central
monitor.
z
In case of DS-LANII network, if the “RR Alarm Source” is other than
Impedance , the RESP waveform will not be transmitted. If the “RR Alarm
Source” is other than GAS , the CO
2
waveform will not be transmitted.
In case of DS-LAN
III
network, refer to the operation manual for the central
monitor.
z
Depending on the central monitor model type, the ST display will be
distorted if the ECG lead (ECG1 or ECG 2) is changed on the DS-7000.
Redrawing the ST display will return the display to normal.
z
On the central monitor, the respiration waveform and RR value based on
the RR/APNEA alarm source selected on the DS-7000 will be displayed.
The RR and APNEA monitored on the central monitor and the DS-7000 will
be the same.
8
-
22
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...