6
-
13
1
Select the pace pulse mask time.
Select from 10ms , 20ms , 40ms depending on the pace
pulse amplitude or presence of fusion beat.
Selecting Auto will automatically mask the pace pulse
depending on the amplitude.
Selecting OFF will set the mask time to 0ms.
●
AC Filter
1
Select ON or OFF for the AC filter.
Selecting ON will set the AC filter which attenuates the AC
noise of 50 to 60Hz.
Selecting OFF will not set the AC filter.
●
ECG Drift Filter
By setting the ECG drift filter ON, only the amplitude with frequency component under 1Hz will be
attenuated to prevent the ECG baseline to drift.
1
Select ON or OFF for the ECG drift filter.
Selecting ON will set the ECG drift filter and controls the
baseline drift.
When the ECG drift filter is set, the patient signal display will
be delayed about 0.5 seconds.
The message, “Drift-F ON” will be displayed on the home
display.
Selecting OFF will not set the ECG drift filter.
Reference
Instead of “Drift-F ON” message, the enlarged time can be displayed depending on the setup.
For procedures, refer to “4. Monitoring Setup –Display Setup–”.
Message
6
ECG
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...