RR Alarm/Apnea Alarm
1
Press the RR Alarm key to display the alarm setup menu.
2
Set ON/OFF of the RR alarm.
ON will generate a RR alarm.
OFF will not generate a RR alarm.
3
Press the RR key, and set the upper and lower limit for the RR alarm.
The set alarm limit will be applied to RR measured from
the impedance respiration waveform, ventilator, or
multigas unit.
The adjustable increment for upper and lower limit
depends on the patient classification.
Adult :
5Bpm
increment
Child/Neonate : 2Bpm increment
Key
Item
Description
▲
Current Value
Indicates the current value.
Lower Alarm
Limit
Range: 5 to 145Bpm (Adult)
2 to 148Bpm (Child/Neonate)
Setting the value to 5Bpm or below (2Bpm or below for
child/neonate) will turn OFF the alarm.
Upper Alarm
Limit
Range: 10 to 150Bpm (Adult)
4 to 150Bpm (Child/Neonate)
Setting the value to 150Bpm or above will turn OFF the
alarm.
Automatic Setup
Automatically sets the upper limit to +20Bpm, and the
lower limit to –20Bpm from the current value.