The “Recorder Error (Central)” message is displayed.
Cause
: The thermal printhead temperature increase or other failure of the central recorder has
occurred.
Solution : Damage to the thermal printhead or other failure can be considered. Contact our service
representative.
The “Recorder Error (8ch.)” message is displayed.
Cause
: The thermal printhead temperature increase or other failure of the 8ch recorder has
occurred.
Solution : Damage to the thermal printhead or other failure can be considered. Contact our service
representative.
Wired Network (DS-LAN
II
/DS-LAN
III
)
The data cannot be displayed on the central monitor.
Cause 1 : The DS-LAN setup is not correct.
Solution : Make sure that the DS-LAN Setup (DS-LAN
II
/DS-LAN
III
) for all bedside monitors and
central monitors in the same network are the same. If the DS-LAN setting is changed,
make sure to restart the system.
Cause 2 : A central monitor which is not compatible with the DS-LAN
III
network is used.
Solution : The following central monitors can not be used with the DS-LAN
III
network.
・
DS-5700
・
DS-5800N/NX/NX
MB
・
DS-7600/7600W with software version V05 and prior
When using these central monitors, all monitors in the same network should be set to
DS-LAN
II
.
Cause 3 : Inappropriate HUB is used.
Solution : Use a repeater HUB recommended by Fukuda Denshi for DS-LANII network and a
switching HUB recommended by Fukuda Denshi for DS-LANIII network.
Cause 4 : The bed ID is duplicated in the same network.
Solution : If bedside monitors with the same bed ID exist in the same network, communication is
not possible. Make sure to set a unique bed ID for each bedside monitor.
Cause 5 : An equipment not specified by Fukuda Denshi is connected to the network.
Solution : Do not connect PC, printer, or other unspecified equipment to the network.
Cause 6 : The DS-LAN cable is not properly connected.
Solution : The DS-LAN connection will be performed by our service representative. Contact our
service representative.
The ECG waveform is not displayed on the central monitor although other waveforms are
displayed.
Cause
: PR is set for “HR/PR”, and BP is set for “PR Source”.
Solution : Set HR for “HR/PR”, or set SpO
2
for “PR Source”.
However, if SpO
2
is not measured while SpO
2
is set for “PR Source”, the setting will
change to BP .
The CO
2
waveform is not displayed on the central monitor although the CO
2
numeric data is
displayed.
Cause 1 : “RR Alarm Source” on the RESP setup menu is set to Impedance .
Cause 2 : “RR Alarm Source” is set to Ventilator .
Solution : Set the “RR Alarm Source” to GAS .
The impedance respiration waveform is not displayed on the central monitor although the RR
numeric data is displayed.
Cause 1 : “RR Alarm Source” on the RESP setup menu is set to GAS .
Trou
ble
sho
oting
10
10
-
23
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...