4
-
66
4
Select the recording factor.
Select the recording factor for alarm recording.
HR will start the alarm recording when a HR or PR alarm is
generated.
Others will start the alarm recording when a numeric alarm
other than HR and PR alarm is generated.
Arrhythmia will start the alarm recording when an
arrhythmia alarm is generated.
5
Select the recording duration.
Select the recording duration from 12sec. , 24sec. .
The recording will automatically stop after the selected time.
If 8ch recorder is selected as the output recorder, Page can
be selected.
The delay time differs depending on the recording duration.
Delay Time
Neonate
Recording
Duration
Adult Child
Numeric Alarm Arrhythmia Alarm
12 sec.
12 sec.
8 sec.
12 sec.
12 sec.
8 sec. for multigas unit alarm
24 sec.
16 sec.
16 sec.
16 sec.
16 sec.
N O T E
6
Select the arrhythmia type.
If arrhythmia is selected for the recording factor, select the arrhythmia type.
Pressing the Arrhythmia Factor key will display the
arrhythmia selection window for alarm recording.
Selected as alarm recording factor.
Not selected as alarm recording factor.
7
Start the alarm recording.
ON will automatically start the recording at alarm occurrence.
If alarm recording is not required, select OFF .
N O T E
The data at the time of alarm occurrence will be recorded.
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...