ECG Waveform Display during Lead-OFF
When a lead–off condition is detected, whether or not to display the waveform for detached lead can
be selected.
1
Set the “ECG Waveform Display during Lead-OFF” on the 3rd page of the ECG setup menu.
ON will display the ECG waveform even during lead-off
condition.
OFF will display the ECG waveform in baseline during
lead-off condition.
●
HR Display during Lead-OFF
When a lead-off condition is detected for the lead used for HR measurement or arrhythmia analysis,
whether or not to display the HR data can be selected.
1
Set the “HR Display during Lead-OFF” on the 3rd page of the ECG setup menu.
ON will display the HR data even during lead-off
condition.
OFF will not display the HR data during lead-off
condition.